Sani-Flow Digital Thermometer Specifications

Pressure Gauges

TOLERANCES:

Compliance: More Info - Compliance 3-A, NEMA 4X, IP67
COMPLIANCE

close 3-A Conforms to 3A 74-05 Sanitary Standards for Sensors and Sensor Fittings and Connections Used on Milk and Milk Products Equipment.

3-A Sanitary Standards, Inc. (3-A SSI) is an independent, not-for-profit corporation dedicated to advancing hygienic equipment design for the food, beverage, and pharmaceutical industries.

NEMA 4 enclosures are used indoors and outdoors to house such items as electrical/ electronic controls, instrumentation systems, and pneumatic, hydraulic and machine tool controls. These enclosures are designed to provide protection against dust, dirt, oil and water. They are suitable for use in areas such as dairies, breweries and similar environments where they are subject to frequent high pressure hosing and generally wet conditions.

IP67 - IP means "Ingress Protection". IP67 enclosures are rated as "dust tight" and protected against the effects of temporary immersion in water.

cse
Accuracy: More Info - Accuracy ± 0.5°F (± 0.3°C)
ACCURACY

close Accuracy is defined as the conformity of an indication to its true value. Accuracy is ± (plus or minus) a stated value across the full range. For example, a thermometer that has an accuracy of ±1/2° would mean that the thermometer can be off no more than ±0.5° anywhere within its range.

cse
Resolution: More Info - Resolution 0.1°F or °C (Full Scale)
RESOLUTION

close Resolution
The resolution of a thermometer is to what fraction of a degree it is possible to make a reading. (0.1°F or °C)

cse
Units: More Info - Units °F or °C - Field Selectable
UNITS

close Switchable

LCD Display Size: More Info - LCD Display Size 1" (2.54 cm)
LCD DISPLAY

close Display The high contrast LCD's numerals are one inch tall (2.54cm).

Refresh Rate: More Info - Refresh Rate Once per Second
REFRESH RATE

close The refresh rate is how often the thermometer takes a reading and displays the result.

cse
Process Temperature Range: More Info - Process Temperature Range -58° to 302°F (-50° to 150°C)
PROCESS TEMPERATURE RANGE

close This is the temperature range which the thermometer is able to measure.

cse
Ambient Operating Limits: More Info - Ambient Operating Limits -40° to 176°F (-40° to 80°C)
AMBIENT OPERATING LIMITS

close This is the temperature range of the environment which the thermometer is able to operate within. If the electronics within the case of the thermometer should go above or below these temperature limits, the thermometer will not operate properly.

cse
Ambient Storage Limits: More Info - Ambient Storage Limits -40° to 185°F (-40° to 85°C)
AMBIENT STORAGE LIMITS

close This is the temperature range of the environment which the thermometer can be stored. If the electronics within the case of the thermometer should go above or below these temperature limits, the thermometer might be damaged.

cse

CONSTRUCTION:

Product Contact Material: 316L Stainless Steel
Product Contact Finish: Ramax = 8 µin (0.20 µm)
Case & Lens Ring Material: 304L Stainless Steel
Lens Material: 1/4" Polycarbonate
Power: 2 Standard AAA Batteries
Battery Life Expectancy: 1 to 3 Years
Approx. Shipping Weight: 2 Pounds (0.9 Kg)

Download a PDF of These Specifications


Configure a Part Number | Specifications | Outline Drawings | Brochure | Battery Replacement & Recalibration